COSYCOT INFANT WARMER
Report
- Report Number
- 9611451-2011-00537
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 3, 2011
- Report Date
- August 5, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- FMT
- PMA / PMN Number
- K971695
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE POWER PCB OF THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE RETURNED POWER PCB OF THE COMPLAINT DEVICE WAS VISUALLY INSPECTED FOR DAMAGE AND WAS INSTALLED ON A WORKING UNIT FOR TESTING. RESULTS: VISUAL INSPECTION REVEALED THAT THE U17 CHIP OF THE POWER PCB WAS DAMAGED. THE BOARD DETAILS INDICATE THAT THE BOARD IS APPROXIMATELY SIX YEARS OLD. THE UNIT WAS UNABLE TO START UP DURING TESTING AS THE POWER FAIL ALARM AND "SEE MANUAL" LED DISPLAYED ONCE WHEN THE UNIT WAS POWERED ON. THE CONTROLLER OF THE TEST UNIT WAS NOT ABLE TO COMPLETE THE START UP SEQUENCE AND THE POWER PCB WAS CONFIRMED TO BE FAULTY. HOWEVER, THE POWER FAIL ALARM CIRCUITRY WAS ABLE TO FUNCTION NORMALLY WHEN THE MAINS ELECTRICAL POWER WAS REMOVED. CONCLUSION: THE FAULT WAS ISOLATED TO THE FAULTY U17 CHIP ON THE POWER PCB THAT PREVENTS THE CONTROLLER FROM STARTING UP. IT IS POSSIBLE THAT THE U17 CHIP WAS DAMAGED BY ELECTRICAL OVER-STRESS, WHICH MAY BE CAUSED BY AN APPLICATION OF EXTREME HIGH VOLTAGE WITH UNLIMITED CURRENT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS TYPE FOR LOT 100525. A REPLACEMENT POWER PCB HAS BEEN PROVIDED TO THE CUSTOMER.
A HOSPITAL IN (B)(6) REPORTED THAT THE POWER FAIL ALARM ON AN IW934JEU COSYCOT INFANT WARMER WAS NOT WORKING. NO PATIENT CONSEQUENCE WAS REPORTED.
A HOSPITAL IN (B)(6) REPORTED THAT THE POWER FAIL ALARM ON AN IW934JEU COSYCOT INFANT WARMER WAS NOT WORKING. NO PATIENT CONSEQUENCE WAS REPORTED.