FDA Adverse Event Malfunction Summary report: N

COSYCOT INFANT WARMER

MDR report key: 2232103 · Received September 1, 2011

Report

Report Number
9611451-2011-00537
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 3, 2011
Report Date
August 5, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
FMT
PMA / PMN Number
K971695
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE POWER PCB OF THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RETURNED POWER PCB OF THE COMPLAINT DEVICE WAS VISUALLY INSPECTED FOR DAMAGE AND WAS INSTALLED ON A WORKING UNIT FOR TESTING. RESULTS: VISUAL INSPECTION REVEALED THAT THE U17 CHIP OF THE POWER PCB WAS DAMAGED. THE BOARD DETAILS INDICATE THAT THE BOARD IS APPROXIMATELY SIX YEARS OLD. THE UNIT WAS UNABLE TO START UP DURING TESTING AS THE POWER FAIL ALARM AND "SEE MANUAL" LED DISPLAYED ONCE WHEN THE UNIT WAS POWERED ON. THE CONTROLLER OF THE TEST UNIT WAS NOT ABLE TO COMPLETE THE START UP SEQUENCE AND THE POWER PCB WAS CONFIRMED TO BE FAULTY. HOWEVER, THE POWER FAIL ALARM CIRCUITRY WAS ABLE TO FUNCTION NORMALLY WHEN THE MAINS ELECTRICAL POWER WAS REMOVED. CONCLUSION: THE FAULT WAS ISOLATED TO THE FAULTY U17 CHIP ON THE POWER PCB THAT PREVENTS THE CONTROLLER FROM STARTING UP. IT IS POSSIBLE THAT THE U17 CHIP WAS DAMAGED BY ELECTRICAL OVER-STRESS, WHICH MAY BE CAUSED BY AN APPLICATION OF EXTREME HIGH VOLTAGE WITH UNLIMITED CURRENT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS TYPE FOR LOT 100525. A REPLACEMENT POWER PCB HAS BEEN PROVIDED TO THE CUSTOMER.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THE POWER FAIL ALARM ON AN IW934JEU COSYCOT INFANT WARMER WAS NOT WORKING. NO PATIENT CONSEQUENCE WAS REPORTED.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THE POWER FAIL ALARM ON AN IW934JEU COSYCOT INFANT WARMER WAS NOT WORKING. NO PATIENT CONSEQUENCE WAS REPORTED.