ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2011-00533
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Report Date
- August 15, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NZ
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT EVAQUA (EXPIRATORY) LIMB OF THE RT340 BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) (B)(4) AND VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT AS THE EVAQUA LIMB WAS TAKEN OUT OF THE PACKAGING, THE FILM OF THE LIMB WAS BREAKING IN DIFFERENT LOCATIONS. CLOSER INSPECTION REVEALED THAT THE FILM, WHICH IS NORMALLY FLEXIBLE, WAS HARD AT THE AREAS WHERE IT SEPARATED. A LOT CHECK WAS NOT PERFORMED AS NO LOT NUMBER WAS PROVIDED. CONCLUSION: THE CAUSE OF THE REPORTED FAULT IS THE UNRAVELLING AND BREAKDOWN OF THE EVAQUA LIMB MATERIAL. ALTHOUGH THE CUSTOMER DID NOT PROVIDE ANY INFORMATION REGARDING THE TYPE OF DRUGS USED, IT IS KNOWN THAT THE USE OF TYLOXAPOL CAN CAUSE THE BREAKDOWN OF THE EVAQUA LIMB. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS PRIOR TO DISTRIBUTION. ANY BREATHING CIRCUIT WHICH DOES NOT PASS THE PRESSURE TEST IS DISCARDED. THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: DO NOT USE MEDICATIONS CONTAINING TYLOXAPOL (SUCH AS TACHOLIQUIN) AS THIS MAY DAMAGE THE TUBING AND LEAD TO A LOSS OF VENTILATION PRESSURE; PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT; SET APPROPRIATE VENTILATOR ALARMS. (B)(4).
A HOSPITAL IN (B)(4) REPORTED THAT A LEAK WAS HEARD FROM THE RT340 BREATHING CIRCUIT AND TWO HOLES WERE OBSERVED ON THE EXPIRATORY LIMB. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |