VENTED AUTOFEED CHAMBER

MDR report key: 2232099 · Received September 1, 2011

Report

Report Number
9611451-2011-00538
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 1, 2011
Report Date
August 3, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). WE HAVE RECENTLY RECEIVED THE COMPLAINT DEVICE AND HAVE NOT BEGUN THE INVESTIGATION PROCESS. WE WILL PROVIDE A FOLLOW-UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) AND VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT THE CONNECTION BETWEEN THE CHAMBER AND THE WATER FEEDSET TUBING WAS BROKEN. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 101006. CONCLUSION: THE FEEDSET TUBE ON THE RETURNED MR290V CHAMBER WAS FOUND BROKEN. THE BREAK WAS ROUGH, SUGGESTING THAT THE DAMAGE MAY HAVE OCCURRED AS A RESULT OF THE FEEDSET TUBE BEING PULLED AWAY FROM THE DOME, POSSIBLY DUE TO BEING CAUGHT OR UNDER TENSION. AS PART OF OUR MANUFACTURING PROCESS, EVERY MR290 CHAMBER UNDERGOES PRESSURE TESTING FOR POTENTIAL LEAKS AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE FEEDSET WAS DAMAGED POST-PRODUCTION. THE MR290 USER INSTRUCTIONS STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARM." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT A TEAR WAS FOUND BETWEEN THE WATER FEEDSET TUBE AND THE DOME OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(4) REPORTED THAT A TEAR WAS FOUND BETWEEN THE WATER FEEDSET TUBE AND THE DOME OF AN MR290 AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 101006

Patients

Seq Age Sex Outcome Treatment
1