FDA Adverse Event Malfunction Summary report: N

OPTIMA MAX, 130K ULTRACENTRIFUGE, 120V

MDR report key: 2232027 · Received September 1, 2011

Report

Report Number
2050012-2011-05103
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 4, 2011
Report Date
August 4, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JQC
PMA / PMN Number
EXEMPT
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) NOTED THAT THE VACUUM PUMP OIL WAS LEAKING INTO THE DIFFERENTIAL PUMP, CAUSING THE BURNING SMELL. FSE FLUSHED THE VACUUM PUMP SEVERAL TIMES, REPLACED THE OIL, OIL FILTER, AND THE VACUUM PUMP MUFFLER ASSEMBLY. THE FSE REPAIRED THE UNIT AND TESTED IT TO CONFIRM PROPER OPERATION WITH NO FURTHER BURNING SMELL. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT A BURNING SMELL FROM THEIR OPTIMA MAX ULTRACENTRIFUGE. NO REPORT OF INJURY HAS BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OPTIMA MAX, 130K ULTRACENTRIFUGE, 120V GENERAL PURPOSE LABORATORY EQUIPMENT LABELED OR PROMOTED FOR A SPECIFIC MEDICAL JQC BECKMAN COULTER INC. OPTIMA MAX N/A

Patients

Seq Age Sex Outcome Treatment
1