FDA Adverse Event
Malfunction
Summary report: N
OPTIMA MAX, 130K ULTRACENTRIFUGE, 120V
MDR report key: 2232027
·
Received September 1, 2011
Report
- Report Number
- 2050012-2011-05103
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 4, 2011
- Report Date
- August 4, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JQC
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BEC FIELD SERVICE ENGINEER (FSE) NOTED THAT THE VACUUM PUMP OIL WAS LEAKING INTO THE DIFFERENTIAL PUMP, CAUSING THE BURNING SMELL. FSE FLUSHED THE VACUUM PUMP SEVERAL TIMES, REPLACED THE OIL, OIL FILTER, AND THE VACUUM PUMP MUFFLER ASSEMBLY. THE FSE REPAIRED THE UNIT AND TESTED IT TO CONFIRM PROPER OPERATION WITH NO FURTHER BURNING SMELL. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT A BURNING SMELL FROM THEIR OPTIMA MAX ULTRACENTRIFUGE. NO REPORT OF INJURY HAS BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OPTIMA MAX, 130K ULTRACENTRIFUGE, 120V | GENERAL PURPOSE LABORATORY EQUIPMENT LABELED OR PROMOTED FOR A SPECIFIC MEDICAL | JQC | BECKMAN COULTER INC. | OPTIMA MAX | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |