FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2232010 · Received September 1, 2011

Report

Report Number
6000001-2011-22243
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
May 27, 2011
Report Date
August 11, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT CONFIRMED THAT THE YELLOW COIL CAP SEPARATED FROM THE COVER. THE SYSTEMIC ROOT CAUSE OF THIS FAILURE MODE WAS DETERMINED TO BE A MANUFACTURING DEFECT; BOTTLE SHOULDER AND LOW ENGAGEMENT AT THE CAP/BOTTLE INTERFACE CAUSED INTERFERENCE BETWEEN THE CAP AND BOTTLE. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION OF THE SEPARATED/CRACKED COIL CAP IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, THIS PRODUCT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR HAD A COIL CAP SEPARATE FROM THE HOUSING DURING INFUSION. THE CONCOMITANT MEDICAL PRODUCTS ARE CURRENTLY UNKNOWN. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 08J070

Patients

Seq Age Sex Outcome Treatment
1