IMMAGE IMMUNOCHEMISTRY SYSTEM

MDR report key: 2231995 · Received September 1, 2011

Report

Report Number
2050012-2011-04696
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
February 21, 2011
Report Date
February 21, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JQX
PMA / PMN Number
K962294
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) REPLACED A DEFECTIVE CIRCULATION FAN IN THE REACTION WHEEL CHAMBER, WHICH WAS THE CAUSE OF THE EVENT. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT A REACTION WHEEL TEMPERATURE ERROR AFTER A PREVENTIVE MAINTENANCE. THE CUSTOMER NOTED BURNING SMELL WHEN CUVETTE COVER WAS LIFTED. THE CUSTOMER SHUT DOWN THE INSTRUMENT. NO INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 IMMAGE IMMUNOCHEMISTRY SYSTEM NEPHELOMETER FOR CLINICAL USE JQX BECKMAN COULTER INC. IMMAGE N/A

Patients

Seq Age Sex Outcome Treatment
1