FDA Adverse Event
Malfunction
Summary report: N
IMMAGE IMMUNOCHEMISTRY SYSTEM
MDR report key: 2231995
·
Received September 1, 2011
Report
- Report Number
- 2050012-2011-04696
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- February 21, 2011
- Report Date
- February 21, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JQX
- PMA / PMN Number
- K962294
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BEC FIELD SERVICE ENGINEER (FSE) REPLACED A DEFECTIVE CIRCULATION FAN IN THE REACTION WHEEL CHAMBER, WHICH WAS THE CAUSE OF THE EVENT. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) TO REPORT A REACTION WHEEL TEMPERATURE ERROR AFTER A PREVENTIVE MAINTENANCE. THE CUSTOMER NOTED BURNING SMELL WHEN CUVETTE COVER WAS LIFTED. THE CUSTOMER SHUT DOWN THE INSTRUMENT. NO INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMAGE IMMUNOCHEMISTRY SYSTEM | NEPHELOMETER FOR CLINICAL USE | JQX | BECKMAN COULTER INC. | IMMAGE | N/A |