INFUSOR
Report
- Report Number
- 6000001-2011-22240
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- May 10, 2011
- Report Date
- August 11, 2011
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- MEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
Narratives
(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, THIS PRODUCT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. VISUAL EXAMINATION OF THE UNIT CONFIRMED THAT THE RED COIL CAP SEPARATED FROM THE COVER. THE SYSTEMIC ROOT CAUSE OF THIS FAILURE MODE WAS DETERMINED TO BE A MANUFACTURING DEFECT; BOTTLE SHOULDER AND LOW ENGAGEMENT AT THE CAP/BOTTLE INTERFACE CAUSED INTERFERENCE BETWEEN THE CAP AND BOTTLE. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION OF THE SEPARATED/CRACKED COIL CAP IS IN PROGRESS THROUGH (B)(4).
BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR HAD A COIL CAP SEPARATE FROM THE HOUSING BEFORE USE. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INFUSOR | PUMP, INFUSION, ELASTOMERIC | MEB | BAXTER HEALTHCARE - IRVINE | 10B073 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |