FDA Adverse Event Injury Summary report: N

REVEAL DX

MDR report key: 2231963 · Received September 1, 2011

Report

Report Number
2647346-2011-01148
Event Type
Injury
Date Received
September 1, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DSI
PMA / PMN Number
K071655
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE LOOP RECORDING WAS HAVING SENSING DIFFICULTIES. THE DEVICE WAS DETECTED FALSE ASYSTOLE AND OVERSENSING OF R-WAVES FOR FAST VENTRICULAR TACHYCARDIA. NEW SOFTWARE WAS LOADED AND THE SENSING MAY BE CHANGED AS WELL. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 REVEAL DX IMPLANTABLE LOOP RECORDER DSI MEDTRONIC MED REL, INC. 9528 ASKU

Patients

Seq Age Sex Outcome Treatment
1 17 YR Required Intervention