FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2231945 · Received September 1, 2011

Report

Report Number
6000001-2011-22208
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
July 13, 2011
Report Date
August 9, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS NOT RETURNED TO BAXTER FOR EVALUATION. THEREFORE, THE REPORTED CONDITION OF A RUPTURED RESERVOIR COULD NOT BE CONFIRMED AND THE ROOT CAUSE COULD NOT BE DETERMINED. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION OF THE BLADDER RUPTURE ISSUE IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE HAD A RESERVOIR RUPTURE DURING PATIENT USE. THE DEVICE WAS FILLED WITH A 150-ML SOLUTION OF 70 MG FABRAZYME IN SALINE. THIS CONDITION INTERRUPTED THERAPY. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 11B078

Patients

Seq Age Sex Outcome Treatment
1 SALINE| FABRAZYME