FDA Adverse Event Malfunction Summary report: N

INFUSOR

MDR report key: 2231938 · Received September 1, 2011

Report

Report Number
6000001-2011-22206
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 8, 2011
Report Date
August 8, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES AND DOES NOT HAVE A 510(K) NUMBER. HOWEVER, THIS PRODUCT IS BEING REPORTED BECAUSE IT IS SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE U.S. THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS NOT CONFIRMED. VISUAL EXAMINATION OF THE SAMPLE SHOWED NO SIGNS OF PHYSICAL ABNORMALITY. A FUNCTIONAL LEAK TEST WAS PERFORMED ON THE PCM. AT THE NORMAL OPERATION SPECIFICATION OF 25 PSI, NO SIGNS OF LEAKAGE WERE OBSERVED ON ANY PART OF THE DEVICE. WHEN THE PRESSURE WAS INCREASED OUT OF OPERATIONAL SPECIFICATION RANGE TO 29 PSI, LEAK WAS NOTED INSIDE THE HOUSING. THEREFORE, THE REPORTED CONDITION OF A LEAK WAS NOT CONFIRMED UNDER NORMAL OPERATIONAL SPECIFICATION OF 25 PSI. NO ROOT CAUSE WAS IDENTIFIED, AS NO EVIDENCE OF PRODUCT MALFUNCTION WAS OBSERVED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INFUSOR PATIENT CONTROL MODULE (PCM) LEAKED DURING PATIENT USE. THE DEVICE WAS BEING USED WITH A SOLUTION OF MORPHINE. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 11C092

Patients

Seq Age Sex Outcome Treatment
1 MORPHINE