ENDOTAK ENDURANCE
Report
- Report Number
- 2124215-2011-15045
- Event Type
- Death
- Date Received
- September 1, 2011
- Date of Event
- June 17, 2011
- Report Date
- August 18, 2011
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, ANALYSIS OF THE LEAD WAS PERFORMED. THREE SEGMENTS OF THE LEAD WERE RETURNED, INCLUDING A 12 CM PORTION OF THE IS-1 TERMINAL LEG, A 10 CM PORTION OF THE PROXIMAL HIGH VOLTAGE TERMINAL LEG, AND A SEGMENT WITH THE DISTAL HIGH VOLTAGE TERMINAL LEG, THE YOKE, AND 36 CM OF THE MIDDLE OF THE LEAD. THE TIP OF THE LEAD WAS NOT RETURNED. INSPECTION OF THE LEAD SEGMENTS NOTED THEY HAD BEEN SEVERED. BLOOD AND BODY FLUID WERE NOTED ON THE TERMINAL PINS, AND INDUCED DAMAGE WAS NOTED ON SEVERAL PARTS OF THE LEAD. INSPECTION OF THE TERMINAL PINS NOTED SETSCREW MARKS CONSISTENT WITH PROPER LEAD CONNECTION. ELECTRICAL TESTS WERE PERFORMED AND ALL THE RETURNED SEGMENTS WERE FOUND TO BE ELECTRICALLY CONTINUOUS. REVIEW OF THE RELATED DEVICE MEMORY FOUND DAILY IMPEDANCE MEASUREMENTS WERE STABLE AND IN RANGE. ANALYSIS OF THE RETURNED SEGMENTS OF THE LEAD COULD NOT REPRODUCE THE REPORTED CLINICAL OBSERVATION OF NOISE AND THE HIGH SHOCK IMPEDANCE OBSERVED IN THE RELATED DEVICE MEMORY.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS LEAD HAD DIED. DIFFICULTY WAS ENCOUNTERED REVIEWING STORED EPISODES AFTER THE DEVICE WAS EXPLANTED. THE RELATED DEVICE WAS RETURNED FOR ANALYSIS. UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, EPISODES WERE SUCCESSFULLY RETRIEVED. REVIEW OF ELECTROGRAMS (EGMS) NOTED THE DEVICE HAD OVERSENSED NOISE, LEADING TO PACING INHIBITION WITH PERIODS OF VENTRICULAR ASYSTOLE FOR THREE TO FIVE SECONDS. REVIEW ALSO NOTED SHOCK IMPEDANCE MEASUREMENTS GREATER THAN 125 OHMS. FURTHER INFORMATION WAS RECEIVED THAT THE DATE OF DEATH WAS SUSPECTED TO BE THE SAME DAY AS THE OUT OF RANGE SHOCK IMPEDANCE MEASUREMENTS AND OVERSENSING. THE PHYSICIAN STATED THE PATIENT DEATH WAS NOT RELATED TO THE NOISE EPISODES. A PORTION OF THE LEAD WAS EXPLANTED AND RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOTAK ENDURANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0154 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Death | 4244| 1861| 0154 |