DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM
Report
- Report Number
- 2517506-2011-00132
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 11, 2011
- Report Date
- August 12, 2011
- Manufacturer
- SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW
- Product Code
- JGS
- PMA / PMN Number
- K051087
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- MEDICAL TECHNOLOGIST
Narratives
ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED SODIUM RESULTS IS UNKNOWN. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.
FALSELY ELEVATED ELECTROLYTE RESULTS FOR SODIUM (NA) WERE OBTAINED ON TWO CONSECUTIVE PATIENT SAMPLES. THE NA RESULTS WERE ABOVE NORMAL REFERENCE RANGES. THE RESULTS WERE REPORTED TO THE PHYSICIAN. THE LAB RECOGNIZED THE TREND AND IMMEDIATELY RERAN THE SAMPLES ON AN ALTERNATE VISTA(R) ANALYZER TO CONFIRM THE RESULTS. THE RESULTS OBTAINED WERE LOWER (NA NOW WITHIN REFERENCE RANGES) AND CORRECTED RESULTS WERE REPORTED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED SODIUM RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIMENSION VISTA® CLINICAL CHEMISTRY SYSTEM | V-LYTE® INTEGRATED MULTISENSOR | JGS | SIEMENS HEALTHCARE DIAGNOSTICS INC - GLASGOW | 1CD859 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |