FDA Adverse Event Malfunction Summary report: N

TAXUS (R) LIBERTÉ (R)

MDR report key: 2231762 · Received September 1, 2011

Report

Report Number
2134265-2011-03716
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 9, 2011
Report Date
August 10, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MFR.: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION STENT DAMAGE OCCURRED. THE 99% STENOSED LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED MID RIGHT CORONARY ARTERY. THE 3.5 X 32MM TAXUS LIBERTE (MR) STENT DELIVERY SYSTEM WAS ADVANCED, BUT WAS UNABLE TO CROSS THE LESION DUE TO THE SEVERELY CALCIFIED LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WERE NO PATIENT COMPLICATIONS AND THE PATIENTS' STATUS WAS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS (R) LIBERTÉ (R) CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - MAPLE GROVE H7493893632350 13967919

Patients

Seq Age Sex Outcome Treatment
1