FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX PTCA DILATATION CATHETER

MDR report key: 2231739 · Received September 1, 2011

Report

Report Number
2134265-2011-03737
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 11, 2011
Report Date
August 12, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(6). (B)(4). DEVICE EVALUATED BY MFR: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE A BALLOON RUPTURE OCCURRED. THE 90% STENOSED TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND MILDLY CALCIFIED MID LEFT ANTERIOR DESCENDING ARTERY. A 12 X 2.75MM NC QUANTUM APEX MR BALLOON WAS INFLATED TO 10ATM ON THE FIRST INFLATION AND RUPTURED ON THE SECOND INFLATION AT 16ATM. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NC QUANTUM APEX PTCA DILATATION CATHETER CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912412270 14193367

Patients

Seq Age Sex Outcome Treatment
1