LEICA M820 F2C
Report
- Report Number
- 3003974370-2011-00018
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- July 25, 2011
- Report Date
- August 23, 2011
- Manufacturer
- LEICA MICROSYSTEMS (SCHWEIZ) AG
- Product Code
- EPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NURSE
Narratives
EXEMPTION NUMBER (B)(4). WHEN THE SURGEON STARTED THE SURGICAL PROCEDURE, THE LIGHT OF THE MICROSCOPE BEGAN BLINKING AND THE FOCUS WAS GOING IN AND OUT. THE SURGEON MOVED THE MICROSCOPE UP AND DOWN TO RESET AND BEGAN AGAIN WITH THE SAME RESULT WITHIN 30 SECONDS OF STARTING. THE LIGHT GOES OFF/ON AND THE MICROSCOPE LOSES FOCUS. THE MICROSCOPE WAS POWERED ON AND OFF TO RESET AGAIN. HOWEVER, THE SAME RESULT WAS OBTAINED (LIGHT GOES OFF/ON AND THE MICROSCOPE LOSES FOCUS). AN ALTERNATE MICROSCOPE WAS BROUGHT IN TO FINISH THE PROCEDURE. NO PATIENT ADVERSE EVENT OCCURRED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A BACK-UP MICROSCOPE. THE NURSE INFORMED THE MANUFACTURER'S REPRESENTATIVE ABOUT THE MALFUNCTION AND ACCORDING TO THE NURSE, THE MICROSCOPE HAS "FOCUS AND CALIBRATION TROUBLE SINCE IT WAS DELIVERED." ON (B)(6) 2011, THE FIELD SERVICE ENGINEER REPAIRED THE MICROSCOPE. ACCORDING TO THE SERVICE REPORT, THE RESET SWITCH AND CONTROL UNIT WERE DEFECTIVE. THE DEFECTIVE PARTS WERE REPLACED AND THE SYSTEM WAS TESTED, CALIBRATED AND PASSED ALL PRIMARY TESTING. THE DEFECTIVE PARTS HAVE BEEN REQUESTED TO BE RETURNED TO THE MANUFACTURER FOR FURTHER INVESTIGATION. ONCE THE RESULTS ARE OBTAINED, A FOLLOW-UP FORM FDA 3500A WILL BE SUBMITTED TO PROVIDE ADDITIONAL INFORMATION.
ON (B)(6) 2011, LEICA MICROSYSTEMS RECEIVED A COMPLAINT STATING THAT THE SURGICAL MICROSCOPE'S LAMP INTERMITTENTLY SHUTS OFF AS WELL AS FOCUS RESETS TO ITS HOME POSITION WHILE THE MICROSCOPE IS IN NORMAL PATIENT WORKING POSITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LEICA M820 F2C | SURGICAL MICROSCOPE | EPT | LEICA MICROSYSTEMS (SCHWEIZ) AG | M820 F2C |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |