COLLEAGUE
Report
- Report Number
- 6000001-2011-22074
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- July 31, 2011
- Report Date
- August 16, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 37:419 WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION BY BAXTER SERVICE PERSONNEL. THE SAMPLE EVALUATION FOUND SEVERAL COMPONENTS THAT CAUSED THE REPORTED CONDITION. THE USER INTERFACE MODULE-PRINTED CIRCUIT BOARD, SOFTWARE, AND THE USER INTERFACE/PUMP HEAD MODULE SIGNAL AND POWER HARNESS WERE ALL DEFECTIVE AND REPLACED TO CORRECT THIS CONDITION. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED SEVEN TIMES BUT NOT FOR THE FOR THE REPORTED CONDITION OF FAILURE CODE 37:419. DURING PREVIOUS SERVICES, THE USER INTERFACE MODULE PRINTED CIRCUIT BOARD WAS REPLACED. A DEVICE HISTORY REVIEW WAS PERFORMED FINDING NO EXCEPTION, NONCONFORMANCE, OR REWORK THAT OCCURRED DURING THE MANUFACTURING OF THE COMPLAINT LOT OR SERIAL NUMBER.
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 37:419. ACCORDING TO THE FACILITY, THIS EVENT OCCURRED UPON POWER UP. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |