FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2231409
·
Received July 18, 2011
Report
- Report Number
- 3004209178-2011-82171
- Event Type
- Malfunction
- Date Received
- July 18, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 30, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED, FAILED THE DISPLACEMENT TEST AND ALARMED MOTOR ERROR DURING REWIND AS A RESULT OF A MOTOR ENCODER SIGNAL BEING OUT OF PHASE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND PROCESS. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT SHE WAS WEARING THE INSULIN PUMP WHILE HAVING AN MRI FOR HER KNEE. THE CUSTOMER STATED THAT SHE WAS UNABLE TO PERFORM THE DISPLACEMENT TEST BECAUSE THE INSULIN PUMP CONTINUED TO ALARM MOTOR ERROR. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 60 YR |