FLOGARD
Report
- Report Number
- 6000001-2011-22052
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- July 31, 2011
- Report Date
- August 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K915522
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH A BROKEN DOOR LATCH WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY THE PUMP'S DOOR BEING CRACKED IN ITS LATCH AREA. THE DOOR ASSEMBLY WAS REPLACED TO CORRECT THE REPORTED CONDITION.
(B)(4).THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPORTED A FLO-GARD INFUSION PUMP WITH A BROKEN DOOR LATCH. ACCORDING TO THE FACILITY, IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THE FACILITY DID NOT HAVE INFORMATION REGARDING PATIENT INVOLVEMENT OR WHETHER THERE HAVE BEEN ANY REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |