LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2011-00289
- Event Type
- Malfunction
- Date Received
- July 26, 2011
- Date of Event
- June 18, 2011
- Report Date
- July 25, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED BELT CONNECTOR) HAS BEEN CONFIRMED. UPON EVALUATION, THE ELECTRODE BELT TRUNK CABLE CONNECTOR WAS DAMAGED. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY DUE TO EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED BELT CABLE CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT. DEVICE EVALUATION OF CHARGER/MODEL SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHARGER SCREEN BLACK) HAS BEEN CONFIRMED. UPON RECEIPT, THE CHARGER/MODEM SCREEN WAS BLACK AND WOULD NOT POWER UP. THE CAUSE FOR THE INABILITY TO POWER UP IS DUE TO A DEFECTIVE POWER SUPPLY. THE ROOT CAUSE FOR THE DEFECTIVE POWER SUPPLY CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE CHARGER/MODEM. THE PATIENT RECEIVED A REPLACEMENT CHARGER/MODEM. DEVICE MANUFACTURE DATE: ELECTRODE BELT SN (B)(4); CHARGER/MODEM SN (B)(4): 05/2010.
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT A DAMAGED BELT CONNECTOR. WHEN A PATIENT SERVICE REPRESENTATIVE (PSR) VISITED THE PATIENT TO EXCHANGE THE ELECTRODE BELT, PSR REPORTED A BLACK SCREEN ON THE CHARGER/MODEM. THE PATIENT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT AND CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR |