FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX
MDR report key: 2231238
·
Received July 22, 2011
Report
- Report Number
- 1218950-2011-02100
- Event Type
- Malfunction
- Date Received
- July 22, 2011
- Report Date
- May 26, 2011
- Manufacturer
- PHILIPS HEALTHCARE - ANDOVER
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE CUSTOMER REPORTED THAT THE EQUIPMENT WAS BEEPING AND IT HAS A PASSWORD AND THE CLIENT IS NOT ABLE TO ACCESS IT. THERE WAS NO PT INVOLVEMENT. THE DEVICE WAS EVALUATED BY A PHILIPS FIELD SERVICE ENGINEER. THE SYMPTOM WAS CLARIFIED AS THE RFH (READY FOR USE) INDICATOR SHOWED A SOLID RED "X" AND A PERIODIC CHIRP, THE CHARGE BUTTON FAILED DURING THE OPCHECK. THE ISSUE WAS LOCALIZED TO A PROCESSOR PCA MALFUNCTION.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE EQUIPMENT WAS BEEPING AND IT HAS A PASSWORD AND THE CLIENT IS NOT ABLE TO ACCESS IT. THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEARTSTART MRX | MKJ | PHILIPS HEALTHCARE - ANDOVER | M3535A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |