FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 3ML LL 200 S/C

MDR report key: 22312173 · Received June 23, 2025

Report

Report Number
1213809-2025-00428
Event Type
Malfunction
Date Received
June 23, 2025
Date of Event
June 5, 2025
Report Date
July 2, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096573
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B.3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. E.1. ADDRESS WAS NOT LOCATED AND IL WAS USED. H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

(B)(4) - SUPPLEMENTAL MDR - FOREIGN MATTER. SINCE NO SAMPLES DISPLAYING THE REPORTED CONDITION WERE RECEIVED A POTENTIAL ROOT CAUSE COULD NOT BE DEFINED AND CORRECTIVE ACTIONS ARE NOT NECESSARY. A PHYSICAL SAMPLE IS REQUIRED FOR A MORE THOROUGH EVALUATION AND POTENTIAL ROOT CAUSE DETERMINATION. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. PLEASE NOTE IT IS POSSIBLE THE FM/ ¿STICKY GEL SUBSTANCE¿ OBSERVED IN THE SYRINGE IS SILICONE. SILICONE IS AN INERT, NON-TOXIC MEDICAL SUBSTANCE USED AS A LUBRICANT FOR DISPOSABLE HYPODERMIC PRODUCTS. IT IS AN INTEGRAL PART OF THE SYRINGE, ENABLING IT TO PERFORM AS REQUIRED IN VARIOUS CLINICAL APPLICATIONS AND DOES NOT PRESENT A SAFETY OR EFFICACY ISSUE NOR DOES IT IMPACT PRODUCT FUNCTION. SILICONE HAS BEEN IN USE IN THIS APPLICATION FOR OVER 20 YEARS, WITH ESTIMATED DISTRIBUTION WELL IN EXCESS OF (B)(4) UNITS. NO REPORTS ARE KNOWN OF ADVERSE CLINICAL EFFECTS ASSOCIATED WITH THESE PRODUCTS AND UNINTENTIONAL DELIVERY OF SILICONE FLUID LUBRICANT. WE WOULD BE VERY INTERESTED IN EXAMINING PRODUCT THAT DOES NOT MEET YOUR EXPECTATIONS AND OUR QUALITY STANDARDS. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. EXAMINATION OF THE PRODUCT INVOLVED MAY PROVIDE CLARIFICATION AS TO THE CAUSE FOR THE REPORTED FAILURE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD SYRINGE 3ML LL 200 S/C HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL #309657 LOT #5083610. THERE WERE NUMEROUS 3ML LUER-LOK TOP SYRINGES, LOT #5083610, FOUND WITH RESIDUE IN THE TIP OF THE SYRINGE 309657.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1941598 BD SYRINGE 3ML LL 200 S/C PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 5083610 30382903096573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown