FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 22311494 · Received June 23, 2025

Report

Report Number
9617229-2025-10378
Event Type
Injury
Date Received
June 23, 2025
Date of Event
May 27, 2025
Report Date
June 23, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628030138
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE BAKER GRADE II, RUPTURE.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE AND CAPSULAR CONTRACTURE WAS RECEIVED ON JUNE 19, 2025, WITH LOT NUMBER 1222176. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: ¿ RUPTURE: OBSERVED BROKEN DEVICE ASSESSED AS SURGICAL DAMAGE AND MISSING PIECE OF SHELL ASSESSED AS INCONCLUSIVE. ¿ CAPSULAR CONTRACTURE: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NO ADDITIONAL OBSERVATIONS PERFORMED ON THE DEVICE. NO FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "CAPSULAR CONTRACTURE BAKER GRADE II AND A RUPTURE." THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE "CAPSULAR CONTRACTURE BAKER GRADE II AND A RUPTURE." THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2063337 INSPIRA SMOOTH COHESIVE SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1222176 10888628030138

Patients

Seq Age Sex Outcome Treatment
1 41 YR Female Required Intervention