FDA Adverse Event Injury Summary report: N

RESTORE PRIME ADVANCED

MDR report key: 2231131 · Received August 26, 2011

Report

Report Number
3004209178-2011-06954
Event Type
Injury
Date Received
August 26, 2011
Date of Event
August 16, 2011
Report Date
August 16, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LGW
PMA / PMN Number
P840001
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

OUT-OF-RANGE IMPEDANCE VALUES WERE REPORTED. IT WAS THOUGHT THAT THE OUT OF RANGE VALUES MIGHT HAVE BEEN DUE TO THE POSSIBILITY THAT ONLY ONE OCTAD AND ONE QUAD WERE IMPLANTED, AS THE REPRESENTATIVE STATED THAT #12-15 WERE OUT OF RANGE. IT WAS REPORTED THAT THE BATTERY LEVEL WAS 2.9 AND THE PT HAD A FEW MORE YEARS REMAINING. ADDITIONAL INFO RECEIVED REPORTED THAT THE HCP WANTED TO REPLACE THE IMPLANTABLE NEUROSTIMULATOR DUE TO THE END OF BATTERY LIFE BASED ON USAGE. THE PT WAS RECEIVING STIMULATION BUT IT WAS NOT COVERING THE PAIN. REPROGRAMMING AND IMPEDANCE CHECKS COULD NOT FIX THE PROBLEM. THE PT RECEIVED A NEW IMPLANTABLE PULSE GENERATOR AND LEADS. THE PT RECEIVED GOOD COVERAGE AND PAIN REDUCTION. ADDITIONAL INFO HAS BEEN REQUESTED, BUT WAS NOT AVAILABLE AS OF THE DATE OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE PRIME ADVANCED LGW MDT PUERTO RICO OPERATIONS CO., JUNCOS 37702 NA

Patients

Seq Age Sex Outcome Treatment
1 63 YR Required Intervention EXTENSION: MODEL 37081, LOT# NJB005057V| LEAD: MODEL 3778, LOT# V022804003| LOT# NKL701369H| EXPLANTED:| IMPLANTED:| EXPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37742, LOT# NJD043038N| LEAD: MODEL 3778, LOT# V022804002| IMPLANTED:| EXTENSION: MODEL 37081, LOT# NJB006008V| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL 37083, LOT# NKC012895N| EXPLANTED:| IMPLANTED:| LEAD: MODEL 3778, LOT# V027479006| EXPLANTED:| IMPLANTED:| EXPLANTED:| IMPLANTED:| EXPLANTED:| PROGRAMMER: MODEL 37742, LOT# NJD046218N| IMPLANTABLE NEURO STIMULATOR: MODEL 37702,