FDA Adverse Event Injury Summary report: N

TAXUS (TM) LIBERTÉ (TM)

MDR report key: 2231104 · Received September 1, 2011

Report

Report Number
2134265-2011-03630
Event Type
Injury
Date Received
September 1, 2011
Date of Event
August 5, 2011
Report Date
August 8, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MANUFACTURER: THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE STENT DISLODGEMENT OCCURRED. ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE 80% STENOSED, 2.75X12-24MM DE NOVO AND ECCENTRIC TARGET LESION WAS LOCATED IN THE MODERATELY TORTUOUS AND NON CALCIFIED RIGHT CORONARY ARTERY (RCA). A 2.75X12MM PROMUS ELEMENT STENT WAS DEPLOYED IN THE DISTAL RCA AND A 2.75X24MM PROMUS ELEMENT STENT WAS DEPLOYED IN THE PROXIMAL RCA. IT WAS NOTED THAT THERE WAS STILL A RESIDUAL LESION JUST PROXIMAL TO THE STENT IN THE DISTAL RCA. A 2.75X12MM TAXUS LIBERTE STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO LESION AND AFTER CROSSING THE PREVIOUSLY PLACED 2.75X24MM PROMUS ELEMENT STENT, THE TAXUS LIBERTE STENT DISLODGED FROM THE SDS IN THE PROXIMAL RCA. A 2.5X9MM NON BSC BALLOON WAS USED TO CRUSH THE DISLODGED STENT AGAINST THE VESSEL WALL. A 2.75X16MM TAXUS LIBERTE STENT WAS THEN DEPLOYED AT 14ATMS, COVERING THE RESIDUAL LESION AND OVERLAPPING THE DISTAL STENT. THE PROCEDURE WAS COMPLETED AT THIS POINT WITH NO PATIENT COMPLICATIONS REPORTED. THE PATIENT'S CURRENT CONDITION IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TAXUS (TM) LIBERTÉ (TM) CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894012270 0013933146

Patients

Seq Age Sex Outcome Treatment
1 57 YR Required Intervention 2.75X12MM PROMUS ELEMENT STENT| 2.75X24MM PROMUS ELEMENT STENT| 0.014 190 CM WHISPER MS SUPPORT GUIDE WIRE| JR4 LAUNCHER GUIDE CATHETER