INRATIO
Report
- Report Number
- 2027969-2011-01578
- Event Type
- Malfunction
- Date Received
- July 19, 2011
- Date of Event
- June 20, 2011
- Report Date
- July 19, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- GJS
- PMA / PMN Number
- K021923
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (B)(6) 2011, INRATIO: 1.7, REFERENCE: 3.3, MEAN: 2.50, CONFIDENCE LIMITS: 1.6 - 3.4. THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND REFERENCE VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO FURTHER INVESTIGATION IS REQUIRED. ANALYSIS OF THE CLIENT'S DATA FROM INRATIO AND REFERENCE TESTS REVEALED THAT TEST RESULT COMPARISON MET ACCURACY CRITERIA. PATIENT'S TEST RESULTS WERE OUTSIDE OF THERAPEUTIC RANGE. ROOT CAUSE COULD NOT BE DETERMINED WITH THE INFORMATION CUSTOMER PROVIDED. NO PRODUCT WAS EXPECTED TO BE RETURNED. RETAINED STRIP TEST RESULT COMPARISON MET ACCURACY CRITERIA. (B)(4). DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLDS MONITORED BY QA FOR CORRECTIVE ACTION (B)(4) NO FURTHER ACTION IS REQUIRED AT THIS TIME. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.
CUSTOMER ALLEGES DISCREPANT RESULTS WITH METER COMPARED TO LAB: DATE: (B)(6) 2011, INRATIO: 1.7, LAB: 3.3. PATIENT'S TARGET THERAPEUTIC RANGE IS 2.0-3.0. LAB RESULTS DRAWN 2-3 HOURS AFTER METER RESULT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | GJS | ALERE SAN DIEGO, INC. | 100071 | 234523 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |