FDA Adverse Event Malfunction Summary report: N

PIN COLLET

MDR report key: 2230996 · Received July 19, 2011

Report

Report Number
1811755-2011-02603
Event Type
Malfunction
Date Received
July 19, 2011
Date of Event
January 8, 2011
Report Date
January 8, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
KIJ
PMA / PMN Number
K972367
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE WIRE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE WIRE WAS STICKING IN THE DEVICE DURING SET UP FOR A PROCEDURE. THERE WAS NO PT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PIN COLLET KIJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK