FDA Adverse Event Malfunction Summary report: N

BD INTIMA-II 24GAX0.75IN PRN SLM

MDR report key: 22309645 · Received June 23, 2025

Report

Report Number
3006948883-2025-00214
Event Type
Malfunction
Date Received
June 23, 2025
Date of Event
May 8, 2025
Report Date
July 4, 2025
Manufacturer
BD SUZHOU (MDS)
Product Code
FOZ
UDI-DI
00382903830282
PMA / PMN Number
UNK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

1. DHR/BHR REVIEW (LOT#4171125): 1) THE PRODUCTS OF THIS BATCH WERE ASSEMBLED AT INTIMA II AUTO LINE 2 IN JUL 2024 AND PACKAGED AT R240 PACKAGE LINE IN JUL 2024. WORK ORDER QUANTITY WAS (B)(4) EA. 2) REVIEW THE IN-PROCESS TEST REPORTS AND OUTGOING TEST REPORTS, AND ALL TEST RESULTS MEET THE PRODUCT SPECIFICATIONS. 3) REVIEW THE PRODUCTION RECORDS WITH NO NONCONFORMANCE, DEVIATION OR REWORK ACTIVITIES. 2. NO DEFECTIVE SAMPLES AND PHOTOS HAVE BEEN RECEIVED FOR THE COMPLAINT. 3. THE RETAINED SAMPLE OF THIS BATCH IS TAKEN FOR 45PSI LEAKAGE TEST, THE TEST IS PASSED, AND NO LEAKAGE IS FOUND AT THE CATHETER. CONCLUSION(S): NO ABNORMALITY IS FOUND ON PROCESS AND RETAINED SAMPLE. BECAUSE DEFECTIVE SAMPLES WERE NOT RETURNED, THE SPECIFIC LEAKAGE SITE AND ABNORMAL STATE OF THE DEFECTIVE SAMPLE CANNOT BE IDENTIFIED, SO THE ROOT CAUSE OF LEAKAGE CANNOT BE DETERMINED. THE PLANT WILL CONTINUE TO TRACK AND MONITOR THE DEFECT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INTIMA-II 24GAX0.75IN PRN SLM LEAKED. THE PATIENT WAS CLEAR AND IN FAIR SPIRITS, PERFORMING INFUSION THERAPY USING AN INTRAVENOUS INDWELLING NEEDLE, AND CHECKING THE STATUS OF THE INDWELLING NEEDLE DURING VENTING REVEALED A LEAK AT THE QYLF NEEDLE HOSE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
837183 BD INTIMA-II 24GAX0.75IN PRN SLM INTRAVASCULAR CATHETER FOZ BD SUZHOU (MDS) 4171125 00382903830282

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown