FDA Adverse Event Malfunction Summary report: N

ARCHITECT HBSAG

MDR report key: 22308926 · Received June 23, 2025

Report

Report Number
3008344661-2025-00087
Event Type
Malfunction
Date Received
June 23, 2025
Date of Event
January 1, 2017
Report Date
August 28, 2025
Manufacturer
ABBOTT IRELAND DIAGNOSTICS DIVISION
Product Code
LOM
UDI-DI
00380740115678
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 06C36 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4P53, WITH 510K/PMA/BLA NUMBER P110029.

Additional Manufacturer Narrative · 0

DATA AND INFORMATION PROVIDED WITH THE ARTICLE WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. DATA AND INFORMATION WITHIN THE ARTICLE WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. A REVIEW OF TRACKING AND TRENDING FOR THE ARCHITECT HBSAG ASSAY DID NOT IDENTIFY AN INCREASE IN COMPLAINT ACTIVITY RELATED TO THE COMPLAINT ISSUE. DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NON-CONFORMANCES, DEVIATIONS, OR POTENTIAL NON-CONFORMANCES WITH THE ASSAY AND THE COMPLAINT ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. IN THIS CASE, THE PATTERN OF SPECIFIC MARKERS IN THIS SAMPLE MOST LIKELY REPRESENTS OCCULT HEPATITIS B INFECTION WITH MUTATIONS BEING A LIKELY CAUSE OF THE OBSERVED NONREACTIVE RESULT. REVIEW OF THE PUBLICATIONS, H. W. REESINK ET AL., ¿OCCULT HEPATITIS B INFECTION IN BLOOD DONORS¿, VOX SANGUINIS (2008) 94 , 153¿166 AND MICHAEL TORBENSON ET AL., ¿OCCULT HEPATITIS B IN HCV+¿, LANCET INFECT DIS 2002; 2: 479¿86, OCCULT HBV INFECTION IS A KNOWN PHENOMENON AND IS DIAGNOSED WHEN A HBV DNA TEST IS POSITIVE BUT HBSAG IS UNDETECTED. THE INFORMATION PRESENTED OF THE VIRUS HAVING COMBINED MULTIPLE MUTATIONS IN THE ANTIGENIC REGION OF THE HBSAG GENE REINFORCES THAT THIS REPRESENTS A KNOWN ESCAPE MUTANT. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WITH THE ARCHITECT HBSAG ASSAY WAS IDENTIFIED.

Additional Manufacturer Narrative · 0

SECTION H MEDICAL DEVICE PROBLEM CODE WAS UPDATED FROM A090803 TO A090810.

Description of Event or Problem · 0

A CONFERENCE SUBMISSION BY L. ZHOU, M P DELCOURT, JM PAWLOTSKY, AND D CANDOTTI, ¿UNUSUAL COMPLEX HEPATITIS B SURFACE PROTEIN VARIANT IN A BLOOD DONOR LEADING TO FALSE-NEGATIVE REACTIVITY WITH A CHEMILUMINESCENT MICROPARTICLE IMMUNOASSAY¿, ISBTABS25-1023, DOCUMENTED FALSE NONREACTIVE ARCHITECT HBSAG RESULTS FOR A BLOOD DONOR. THE CONFERENCE SUBMISSION AND PRESENTATION NOTED A 63 YEAR OLD MALE BLOOD DONOR THAT WAS UNVACCINATED FOR HBV WHO HAD A POSITIVE VIRAL LOAD OF 16 IU/ML AND TESTED NONREACTIVE WITH THE ARCHITECT HBSAG AND PRISM HBSAG IN 2017. ADDITIONAL TESTING WITH ENZYMATIC IMMUNOASSAYS (EIA) GENERATED REACTIVE RESULTS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

A CONFERENCE SUBMISSION BY L. ZHOU, M P DELCOURT, JM PAWLOTSKY, AND D CANDOTTI, ¿UNUSUAL COMPLEX HEPATITIS B SURFACE PROTEIN VARIANT IN A BLOOD DONOR LEADING TO FALSE-NEGATIVE REACTIVITY WITH A CHEMILUMINESCENT MICROPARTICLE IMMUNOASSAY¿, ISBTABS25-1023, DOCUMENTED FALSE NONREACTIVE ARCHITECT HBSAG RESULTS FOR A BLOOD DONOR. THE CONFERENCE SUBMISSION AND PRESENTATION NOTED A 63 YEAR OLD MALE BLOOD DONOR THAT WAS UNVACCINATED FOR HBV WHO HAD A POSITIVE VIRAL LOAD OF 16 IU/ML AND TESTED NONREACTIVE WITH THE ARCHITECT HBSAG AND PRISM HBSAG IN 2017. ADDITIONAL TESTING WITH ENZYMATIC IMMUNOASSAYS (EIA) GENERATED REACTIVE RESULTS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

A CONFERENCE SUBMISSION BY L. ZHOU, M P DELCOURT, JM PAWLOTSKY, AND D CANDOTTI, ¿UNUSUAL COMPLEX HEPATITIS B SURFACE PROTEIN VARIANT IN A BLOOD DONOR LEADING TO FALSE-NEGATIVE REACTIVITY WITH A CHEMILUMINESCENT MICROPARTICLE IMMUNOASSAY¿, ISBTABS25-1023, DOCUMENTED FALSE NONREACTIVE ARCHITECT HBSAG RESULTS FOR A BLOOD DONOR. THE CONFERENCE SUBMISSION AND PRESENTATION NOTED A 63 YEAR OLD MALE BLOOD DONOR THAT WAS UNVACCINATED FOR HBV WHO HAD A POSITIVE VIRAL LOAD OF 16 IU/ML AND TESTED NONREACTIVE WITH THE ARCHITECT HBSAG AND PRISM HBSAG IN 2017. ADDITIONAL TESTING WITH ENZYMATIC IMMUNOASSAYS (EIA) GENERATED REACTIVE RESULTS. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
927733 ARCHITECT HBSAG TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) LOM ABBOTT IRELAND DIAGNOSTICS DIVISION 00380740115678

Patients

Seq Age Sex Outcome Treatment
1 63 YR Male ARC I2K PROC MOD, 03M74-01, 116UNKNOWND11.| ARC I2K PROC MOD, 03M74-01, 116UNKNOWND11.| ARC I2K PROC MOD, 03M74-01, 116UNKNOWND11.