FDA Adverse Event Injury Summary report: N

SAFIRE BI-DIRECTIONAL CATH., 7F, LARGE SWEEP, 4MM

MDR report key: 2230842 · Received August 26, 2011

Report

Report Number
3005188751-2011-00149
Event Type
Injury
Date Received
August 26, 2011
Date of Event
August 6, 2011
Report Date
August 6, 2011
Manufacturer
ST. JUDE MEDICAL, AF DIVISION
Product Code
LPB
PMA / PMN Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE CATHETER WAS NOT RETURNED FOR EVAL. HOWEVER, REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE DEVICE MET MFG REQUIREMENTS PRIOR TO SHIPMENT. THE ROOT CAUSE CLASSIFICATION FOR THE REPORTED AORTIC DISSECTION IS CONSISTENT WITH ANTICIPATED PROCEDURAL COMPLICATIONS AS THIS EVENT IS A KNOWN PHYSIOLOGICAL EFFECT OF THE PROCEDURE AND IS NOTED WITHIN THE IFU.

Description of Event or Problem · 1

IT WAS REPORTED DURING A LATERAL ACCESSORY PATHWAY CARDIAC ABLATION PROCEDURE THE PT DEVELOPED AN AORTIC DISSECTION. WHILE USING THE SAFIRE BI-DIRECTIONAL ABLATION CATHETER FOR MAPPING AND ABLATING THE LEFT ATRIA VIA RETROGRADE APPROACH, THE PHYSICIAN NOTED THE CATHETER LOOP WOULD NOT OPEN AND THE DEVICE WAS DIFFICULT TO MANEUVER. THE PT COMPLAINED OF ABDOMINAL PAIN DURING THE PROCEDURE BUT THE PT WAS HEMODYNAMICALLY STABLE WITH AN ARTERIAL BLOOD PRESSURE OF 150/70, OXYGEN SATURATION OF 95% ON ROOM AIR WHICH ELEVATED TO 99% WITH OXYGEN THERAPY. THE PT SYMPTOMS PERSISTED AND SELOKEN IV WAS ADMINISTERED. A CT SCAN CONFIRMED AN AORTIC DISSECTION AND THE PT WAS TRANSFERRED EMERGENTLY TO A NEARBY FACILITY FOR SURGERY WITH THE CATHETERS IN PLACE. THE PT UNDERWENT SURGICAL REPAIR AND PROSTHESIS OF THE AORTIC ARCH. CURRENT CONDITION OF THE PT IS IMPROVING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SAFIRE BI-DIRECTIONAL CATH., 7F, LARGE SWEEP, 4MM SAFIRE CATHETER, 7F, LRG LPB ST. JUDE MEDICAL, AF DIVISION 402820 3355188

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention NAVX PATCHES, LOT UNK