FDA Adverse Event Injury Summary report: N

KIT IMPLANTABLE SLIM TIP LEAD, 50CM

MDR report key: 22307868 · Received June 23, 2025

Report

Report Number
1627487-2025-03058
Event Type
Injury
Date Received
June 23, 2025
Date of Event
June 2, 2025
Report Date
July 28, 2025
Manufacturer
ABBOTT MEDICAL
Product Code
PMP
UDI-DI
05415067027153
PMA / PMN Number
P150004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT IS ESTIMATED. THE ALLEGATION IS AGAINST 1 OF 2 LEADS; HOWEVER, IT IS UNKNOWN WHICH LEAD, THEREFORE, ALL POTENTIAL COMPONENTS ARE BEING LISTED. ADDITIONAL COMPONENTS POTENTIALLY INVOLVED IN THE EVENT INCLUDE: COMMON DEVICE NAME: DRG LEAD, MODEL: MN10450-50A, UDI: (B)(4), SERIAL: (B)(6), BATCH: 9042817.

Additional Manufacturer Narrative · 0

THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED INCIDENT COULD NOT BE CONCLUSIVELY DETERMINED.

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT'S SYSTEM WAS UNABLE TO ENTER MRI MODE. TROUBLESHOOTING REVEALED HIGH IMPEDANCES ON ONE LEAD. AS A RESULT, SURGICAL INTERVENTION MAY BE UNDERTAKEN TO ADDRESS THE ISSUE. IT IS UNKNOWN WHICH LEAD IS LIABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
970385 KIT IMPLANTABLE SLIM TIP LEAD, 50CM DRG LEAD PMP ABBOTT MEDICAL MN10450-50A 8904026 05415067027153

Patients

Seq Age Sex Outcome Treatment
1 36 YR Male Other DRG IPG| DRG LEAD