FDA Adverse Event Injury Summary report: N

ACHIEVA 1.5T NEW

MDR report key: 2230732 · Received August 17, 2011

Report

Report Number
3003768277-2011-00506
Event Type
Injury
Date Received
August 17, 2011
Date of Event
July 31, 2011
Report Date
August 1, 2011
Manufacturer
PHILIPS HEALTHCARE
Product Code
LNH
PMA / PMN Number
K063559
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) (EVAL METHOD, RESULTS, CONCLUSIONS) - THE INVESTIGATION IS STILL ONGOING ON THIS EVENT. WHEN THE INVESTIGATION IS COMPLETED A FOLLOW-UP REPORT WILL BE SENT TO THE FDA.

Description of Event or Problem · 1

THE PATIENT WAS SCANNED IN A HEAD FIRST SUPINE POSITION WITH THE SPINE 15CH COIL, WHILE A SENSE HEAD COIL WAS ALSO ON THE TABLE, NOT CONNECTED. THE ARM OF THE PATIENT MADE CONTACT WITH THE CABLE OF THE SENSE HEAD COIL DURING SCANNING AND AFTERWARDS A 2ND DEGREE BURN OF APPROXIMATELY 1X3 CM WAS FOUND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACHIEVA 1.5T NEW LNH PHILIPS HEALTHCARE 781296

Patients

Seq Age Sex Outcome Treatment
1 39 YR Other