FDA Adverse Event Malfunction Summary report: N

OFFSET CUP IMPACTOR - M8X1 INTERFACE

MDR report key: 22307005 · Received June 23, 2025

Report

Report Number
3013223655-2025-00001
Event Type
Malfunction
Date Received
June 23, 2025
Date of Event
June 11, 2025
Report Date
June 23, 2025
Manufacturer
INCIPIO DEVICES SA
Product Code
HWA
UDI-DI
07630055302458
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

PHOTOGRAPHS PROVIDED WITH THE EVENT REPORT CONFIRM THE PRODUCT FAILURE. THE STRIKE PLATE HAS FRACTURED AT THE ALIGNMENT GUIDE GROOVE. THE REPORTED ISSUE HAS BEEN IDENTIFIED AS AN ANTICIPATED END OF LIFE FAILURE IN THE PRODUCT RISK MANAGEMENT FILE. THE RISKS ASSOCIATED WITH THE STRIKE PLATE FRACTURE HAVE BEEN REDUCED AS A FAR AS POSSIBLE AND RESIDUAL RISK IS LOW AND ACCEPTABLE. OCCURRENCE LEVELS REMAIN ACCEPTABLE. THE FAILURE WILL CONTINUE TO BE TRACKED AS PART OF ROUTINE TREND MONITORING ACTIVITIES. THE CAUSE OF THE ISSUE IS FATIGUE FOLLOWING MANY YEARS OF USE AND REUSE. NO CORRECTIVE ACTION IS CONSIDERED NECESSARY AT THIS TIME.

Description of Event or Problem · 0

THE BACK PART OF THE IMPACTOR BROKE WHILE THE CUP WAS BEING IMPACTED. SINCE THE IMPACTION WAS ALMOST COMPLETE, NO BACKUP INSTRUMENTS WERE USED AND THE SURGERY WAS COMPLETED SUCCESSFULLY. EVENT OCCURED IN JAPAN. PRODUCT IS ALSO MARKETED IN THE US.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
927581 OFFSET CUP IMPACTOR - M8X1 INTERFACE OFFSET CUP IMPACTOR HWA INCIPIO DEVICES SA 01.15.10.0556 20J10444 07630055302458

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown