FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2230669 · Received September 1, 2011

Report

Report Number
1644487-2011-02038
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
June 18, 2009
Report Date
August 25, 2011
Manufacturer
CYBERONICS INC
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

MANUFACTURER REVIEW OF A VNS PATIENT'S PROGRAMMING HISTORY REVEALED A FAULTED SYSTEMS DIAGNOSTICS TEST OCCURRED ON (B)(6) 2009 WHICH CHANGED THE VNS SETTINGS TO 1MA/20HZ/500 PULSEWIDTH/30 SECONDS ON/60 MINUTES OFF. THE SETTINGS WERE NOT CORRECTED UNTIL (B)(6) 2009.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS INC MODEL 250 NI

Patients

Seq Age Sex Outcome Treatment
1 57 YR