FDA Adverse Event Injury Summary report: N

TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO

MDR report key: 2230648 · Received August 12, 2011

Report

Report Number
2953200-2011-01427
Event Type
Injury
Date Received
August 12, 2011
Date of Event
July 13, 2011
Report Date
July 13, 2011
Manufacturer
MEDTRONIC CARDIOVASCULAR
Product Code
MIH
PMA / PMN Number
P070027
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). RESULTS OF EVAL: (ENDOLEAK), (DISEASE PROGRESSION, SHORT AND DILATED AORTIC NECK). CONCLUSION: (DISEASE PROGRESSION, SHORT AND DILATED AORTIC NECK).

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PT FOR THE ENDOVASCULAR TREATMENT OF A 5 CM IN DIAMETER ABDOMINAL AORTIC ANEURYSM APPROX 8.5 YRS AGO. THE AORTIC NECK WAS SHORT AND CONICAL IN SHAPE. IT WAS REPORTED THAT AT A CT PERFORMED APPROX 50 MONTHS POST-IMPLANT, IT WAS NOTED THAT THERE MAY HAVE BEEN A TYPE I ENDOLEAK (MFR REPORT # 2953200-2007-00394). THE ANEURYSM WAS 5.2 CM IN DIAMETER, AND THERE WAS EXTENSIVE MURAL THROMBUS AND SOME CONTRAST WITHIN THE ANEURYSM SAC. THE ILIAC LIMBS WERE PATENT. APPROX 7 YRS POST-IMPLANT, THE PHYSICIAN ELECTED TO INTERVENE BY FIRST PLACING A RENAL STENT VIA THE ARM AND THEN IMPLANTING A TALENT CONVERTER AND A TALENT AORTIC CUFF (MFR REPORT # 2953200-2011-01427) OVER THE RENAL STENT TO ACHIEVE A PROXIMAL SEAL SUCCESSFULLY, FOLLOWED BY PERFORMING A FEMORAL-FEMORAL BYPASS. APPROX 1 MONTH AGO, IT WAS NOTED THAT THERE WAS A PROXIMAL TYPE I ENDOLEAK AND THAT THE ANEURYSM WAS 7.2 CM IN DIAMETER. THE AORTIC NECK WAS LESS THAN 1 CM LONG AND 25 - 26.5 MM IN DIAMETER. THE ENDOLEAK WAS ATTRIBUTED TO THE SHORT AND DILATED AORTIC NECK AND DISEASE PROGRESSION. THE PT IS NOT A SURGICAL CANDIDATE. THE PHYSICIAN ELECTED TO INTERVENE BY PLACING A CUFF PROXIMALLY, WHICH SUCCESSFULLY RESOLVED THE ENDOLEAK. NO ADD'L CLINICAL SEQUELAE WERE REPORTED, AND THE PT IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TALENT ABDOMINAL STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC CARDIOVASCULAR NA V00231691

Patients

Seq Age Sex Outcome Treatment
1 84 YR Required Intervention