FDA Adverse Event Injury Summary report: N

UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM

MDR report key: 2230646 · Received August 12, 2011

Report

Report Number
2122870-2011-02813
Event Type
Injury
Date Received
August 12, 2011
Date of Event
July 31, 2008
Report Date
July 31, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
MMI
PMA / PMN Number
K023764
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS NOT DISPATCHED AS THE CUSTOMER DID NOT QUESTION SYSTEM PERFORMANCE. CREATINE KINASE-MB (CK-MB) WAS TESTED AND PRODUCED A RESULT ABOVE THE REFERENCE RANGE, BUT THE CUSTOMER DID NOT QUESTION THE RESULT. TESTING AT BECKMAN COULTER CUSTOMER PRODUCT LINE SUPPORT (CPLS) LABORATORY CONFIRMED THE EXISTENCE OF A HETEROPHILE INTERFERENCE FOR THE SAMPLES REC'D FROM THE CUSTOMER. THIS INTERFERING SUBSTANCE CAUSES FALSELY ELEVATED PT RESULTS. PRODUCT LABELING: FOR ASSAYS EMPLOYING ANTIBODIES, THE POSSIBILITY EXISTS FOR INTERFERENCE BY HETEROPHILE ANTIBODIES IN THE PT SAMPLE. PTS WHO HAVE BEEN REGULARLY EXPOSED TO ANIMALS OR HAVE REC'D IMMUNOTHERAPY OR DIAGNOSTIC PROCEDURES UTILIZING IMMUNOGLOBULINS OR IMMUNOGLOBULIN FRAGMENTS MAY PRODUCE ANTIBODIES, E.G. HUMAN ANTI-MOUSE ANTIBODIES (HAMA), THAT INTERFERE WITH IMMUNOASSAYS. ADDITIONALLY, OTHER HETEROPHILE ANTIBODIES SUCH AS HUMAN ANTI-GOAT ANTIBODIES MAY BE PRESENT IN PT SAMPLES. SUCH INTERFERING ANTIBODIES MAY CAUSE ERRONEOUS RESULTS. CAREFULLY EVALUATE THE RESULTS OF PTS SUSPECTED OF HAVING THESE ANTIBODIES. THE ACCESS ACCUTNI AND ACCESS CK-MB RESULTS SHOULD BE INTERPRETED IN LIGHT OF THE TOTAL CLINICAL PRESENTATION OF THE PT, INCLUDING: SYMPTOMS, CLINICAL HISTORY, CLINICAL EXAMINATION, ELECTROCARDIOGRAM (ECG), DATA FROM ADD'L TESTS, AND OTHER APPROPRIATE INFO. MEDICAL DECISIONS SHOULD NOT BE BASED ON A SINGLE ACCUTNI DETERMINATION AT ONE TIME POINT. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF PRODUCT COMPLAINTS CONDUCTED FROM (B)(6) 2008 THROUGH (B)(6) 2010 FOR ADD'L REPORTABLE EVENTS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ELEVATED TROPONIN I (ACCUTNI) RESULTS, ABOVE THE ACUTE MYOCARDIAL INFARCTION (AMI) CUTOFF, FOR ONE PT INVOLVING UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM. THE ELEVATED RESULTS WERE RELEASED OUT OF THE LABORATORY. THE PT REC'D CARDIAC CATHETERIZATION AND HAD GYNECOLOGY SURGERY DELAYED DUE TO THE ELEVATED RESULTS. THE CUSTOMER CONFIRMED THERE WAS NO PT INJURY. THE CUSTOMER SUPPLIED BECKMAN COULTER, INC. WITH THE PT SAMPLE FOR FURTHER ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM MMI BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Other| R ACCESS ACCUTNI| UNICEL DXI ACCESS IMMUNOASSAY SYSTEM (MODEL UNK)