ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-06412
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Report Date
- August 5, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: MOISTURE WAS OBSERVED BEHIND THE PUMP DISPLAY LENS. A LEAK TEST WAS PERFORMED AND THE DISPLAY LENS WAS FOUND TO BE LEAKING. THE PUMP WAS UNABLE TO POWER ON. THE PUMP WAS OPENED AND MOISTURE DAMAGE WAS OBSERVED INSIDE THE PUMP.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
THE PATIENT'S MOTHER REPORTED THAT THE PATIENT RECEIVED AN OCCLUSION ALARM APPROXIMATELY 2 HOURS PRIOR, THEY DECIDED TO CHANGE THE BATTERY AND NOW THE PUMP HAS NO POWER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR |