HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2230407 · Received September 1, 2011

Report

Report Number
1423500-2011-11518
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 11, 2011
Report Date
August 11, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS DISCARDED BY THE CUSTOMER. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT WAS CONFIRMED. THE ROOT CAUSE WAS IDENTIFIED TO BE FROM AIR IN THE PATIENT LINE DESCRIBED BY THE HOME PATIENT. THIS ISSUE IS BEING INVESTIGATED UNDER A CAPA.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER TECHNICAL SERVICES (BTS) REGARDING ASSISTANCE WITH A LOW DRAIN VOLUME ALARM DURING THE INITIAL DRAIN ON THE HOMECHOICE MACHINE (HC). THE HOME PATIENT (HP) STATED THERE WAS AIR THROUGHOUT ALL OF HIS LINES INCLUDING THE PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP TO START OVER WITH NEW SUPPLIES. THE NURSE WAS CONTACTED ON (B)(6) 2011. THE NURSE STATED THAT THE HP DID NOT DEVELOP ANY ADVERSE REACTIONS OR REQUIRE ANY MEDICAL INTERVENTION. THE NURSE STATED THE HP IS CURRENTLY PERFORMING THERAPY WITHOUT ANY COMPLICATIONS. THE NURSE STATED THAT THIS WAS AN ISOLATED EVENT. THERE WAS PATIENT INVOLVEMENT; HOWEVER, THERE WAS NO INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 60 YR HOMECHOICE