HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-11507
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 14, 2011
- Report Date
- August 14, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A BATCH REVIEW WILL NOT BE CONDUCTED.
(B)(4). THIS COMPLAINT WAS FOR A CONNECTION ISSUE AND WAS CONFIRMED BASED ON INFORMATION PROVIDED BY THE HOME PATIENT (HP) STATING THAT THEY DID NOT CONNECT THEIR FINAL BAG ALL THE WAY AND SOME OF THE FLUID LEAKED OUT. THE LABEL REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR THE USE ERROR IN THIS INCIDENT. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
DURING ASSISTANCE WITH SET UP ON THE HOME CHOICE (HC), DURING USE; THE PATIENT REVEALED THEY HAD CONNECTED THE BAGS INCORRECTLY. THEY STATED ONE OF THE BAGS WAS NOT CONNECTED SECURELY AND SOME OF THE FLUID STARTED TO LEAK OUT. THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE PATIENT WITH STARTING OVER. DURING A FOLLOW-UP WITH THE HOME PATIENT (HP) REGARDING THE REPORTED PROBLEM, THE HP STATED THAT THEY DID START OVER WITH NEW SUPPLIES, AND DOUBLE CHECKED THE CONNECTION. THE HP STATED THERAPY WENT FINE AFTERWARDS, AND HAS BEEN GOING FINE SINCE. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | HOMECHOICE |