ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2011-06409
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Report Date
- August 4, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE KEYPAD BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS.
THE PATIENT REPORTED THAT THE OK AND DOWN ARROW BUTTONS WERE FUNCTIONING INTERMITTENTLY. THE PATIENT REPORTED THAT WHILE ATTEMPTING TO PROGRAM A BOLUS, THE OK BUTTON WOULD NOT WORK. THE PATIENT CONFIRMED THAT THERE WAS NO TRAUMA TO THE PUMP, THE PUMP WAS NOT EXPOSED TO WATER, AND THE KEYPAD WAS INTACT. THE PATIENT REPORTEDLY WORE THE PUMP IN A PUMP CASE ATTACHED TO A BELT AND DID NOT CLEAN THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR |