HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1423500-2011-11506
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 10, 2011
- Report Date
- August 10, 2011
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K923065
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE REPORTED ISSUE OF AIR IN TUBING (WITHOUT ALARM) WAS NOT CONFIRMED. THE CAUSE WAS UNDETERMINED. A LABELING REVIEW FOUND THE PATIENT AT HOME GUIDE TO BE ADEQUATE FOR POTENTIAL USE/USER ERROR RELATED TO THIS INCIDENT.
(B)(4). THE SAMPLE IS NOT AVAILABLE AND THE LOT NUMBER IS UNKNOWN. SINCE THE LOT NUMBER IS UNKNOWN, NO BATCH REVIEW WILL BE PERFORMED. A FOLLOW-UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS OBTAINED.
DURING TROUBLESHOOTING FOR AN ALARM ON THE HOMECHOICE MACHINE, THE HOME PATIENT (HP) REPORTED TO BAXTER TECHNICAL REPRESENTATIVE (TSR) THAT THERE WAS AIR IN THE PATIENT LINE. THE HP STATED SHE HAD AIR IN HER PATIENT LINE EVERY NIGHT AND HAD SPOKEN WITH HER DOCTOR. THE TSR GAVE RECOMMENDATIONS OF ENDING THERAPY AND STARTING OVER WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | HOMECHOICE |