FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2230309 · Received September 1, 2011

Report

Report Number
2050012-2011-04755
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 1, 2011
Report Date
August 1, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 AND INSPECTED THE HYDRO CANISTERS AND MOISTURE SEPARATORS WITH NO ISSUES FOUND. THE FSE ALSO ENABLED THE CHEMISTRY CARTRIDGE (CC) SIDE AND PERFORMED A STOP/HOME CYCLE. LOW/HIGH PRESSURES WERE ADJUSTED INTO THEIR RANGES. THE FSE ALSO PRIMED THE INSTRUMENT WITH NO ERRORS. QC WAS TESTED BY THE CUSTOMER FOR VERIFICATION; NO ISSUES WERE NOTED BY FSE. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES PRIOR TO RETURNING THE INSTRUMENT INTO SERVICE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A HYDRO ERROR OCCURRENCE AND A LEAK THAT WAS OBSERVED IN THE HYDRO AREA IN THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE CUSTOMER INDICATED THAT WASTE HAD BACKED UP INTO THE PRIMARY VACUUM ACCUMULATOR. CUSTOMER STATED THEY PULLED THE VACUUM ACCUMULATOR AND THE WASTE EXIT SUMP. FLUID WAS FOUND IN THE DRYERS. NO INJURY WAS REPORTED IN THIS EVENT

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1