COULTER AC*T DIFF 2 ANALYZER
Report
- Report Number
- 1061932-2011-01333
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 1, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K990352
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(6) 2011 FOR THIS EVENT AND INDICATED THAT THE BATH WAS OVERFLOWING DUE TO A CRIMPED WASTE LINE. THE FSE REPLACED THE BLOODY VACUUM ISOLATION CHAMBER (VIC) AND VERIFIED REPAIRS PER ESTABLISHED PROCEDURES. ROOT CAUSE FOR THE LEAK IS A CRIMPED WASTE LINE THAT DID NOT ALLOW THE WBC AND RBC BATHS TO DRAIN PROPERLY. PER LABELING, BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. BEC IDENTIFIER FOR THIS REPORT IS (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REPORTING A POOL OF BIOHAZARD FLUID UNDERNEATH THE COULTER AC*T DIFF 2 ANALYZER. WHILE CYCLING PATIENT SAMPLES A VACUUM FAILURE WAS DISPLAYED. THE CUSTOMER INSPECTED THE INSTRUMENT AND FOUND THE RED BLOOD COUNT (RBC) AND WHITE BLOOD COUNT (WBC) BATHS HAD OVERFLOWED. THE USER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF LAB COAT, GLOVES AND EYE PROTECTION AT THE TIME OF THE INCIDENT. THERE WAS NO REPORT OF ANY PATIENT RESULTS BEING AFFECTED BY THIS INCIDENT. MEDICAL ATTENTION WAS NOT SOUGHT. THERE WAS NO REPORT OF EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER AC*T DIFF 2 ANALYZER | AUTOMATED DIFFERENTIAL CELL COUNTER | GKZ | BECKMAN COULTER, INC. | ACT DIFF 2 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |