FDA Adverse Event Malfunction Summary report: N

BECKMAN COULTER¿ COULTER® LH 750 HEMATOLOGY

MDR report key: 2230305 · Received September 1, 2011

Report

Report Number
1061932-2011-01329
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
July 31, 2011
Report Date
August 1, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K011342
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SAMPLE COLLECTION AND STORAGE INFORMATION WAS NOT PROVIDED. THE INSTRUMENT IS CURRENTLY WITHIN QC SPECIFICATIONS WITH RESPECT TO CONTROLS (ACCURACY) AND REPRODUCIBILITY (PRECISION). CONTROLS WERE RUN BEFORE AND AFTER THIS EVENT AND WERE WITHIN ACCEPTABLE RANGES. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND REPLACED THE BLOOD SAMPLING VALVE (BSV - LINES 207). THE FSE ALSO REPLACED THE RETIC HEATER. THE FSE VERIFIED INSTRUMENT OPERATION PRIOR TO RETURNING IT INTO SERVICE. THE ROOT CAUSE FOR THIS EVENT MAY BE ATTRIBUTED TO THE HARDWARE THAT WAS REPLACED DURING SUBSEQUENT INSTRUMENT SERVICING FOR THIS EVENT.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) IN REGARDS TO OBTAINING ERRONEOUS COMPLETE BLOOD COUNT (CBC) RESULTS (ALL PARAMETERS) THAT WERE GENERATED BY THE BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER FOR TWO (2) PATIENTS. BOTH PATIENT SAMPLES HAD NO INSTRUMENT FLAGS GENERATED ON THE ERRONEOUS RESULTS. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. PER THE CUSTOMER, THEY NOTICED THAT CONTROL RECOVERY DROPPING AFTER THREE PIERCING AND PERFORMED REPRODUCIBILITY TESTING, WHICH FAILED. THE RESULTS PROVIDED BY THE CUSTOMER ARE SHOWN IN THE ATTACHMENT SECTION OF THIS REPORT. THERE WAS NO DEATH, INJURY, OR AFFECT TO PATIENT TREATMENT REPORTED FOR THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BECKMAN COULTER¿ COULTER® LH 750 HEMATOLOGY COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER, INC. LH 750 N/A

Patients

Seq Age Sex Outcome Treatment
1