FDA Adverse Event Malfunction Summary report: N

ALTIUS SYSTEM

MDR report key: 22302807 · Received June 20, 2025

Report

Report Number
3026283936-2025-10005
Event Type
Malfunction
Date Received
June 20, 2025
Date of Event
January 1, 2025
Report Date
June 19, 2025
Manufacturer
NEUROS MEDICAL INC.
Product Code
LGW
PMA / PMN Number
P230020
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ALTHOUGH THE PRODUCT WILL NOT BE RETURNED AT THIS TIME FOR INVESTIGATION, IMPEDANCE MEASUREMENTS, REPORTED PATIENT SYMPTOMS AND HISTORICAL DEVICE KNOWLEDGE SUGGEST THAT THE INTEGRITY OF THE NERVE CUFF ELECTRODE (LEAD) MAY HAVE BEEN COMPROMISED DUE TO THE FALL CAUSING INTERMITTENT CONTINUITY AND A SHOCKING SENSATION. THE SHOCKING SEQUELA HAS SUBSEQUENTLY RESOLVED. THE NERVE CUFF ELECTRODE (LEAD) INVOLVED IN THIS EVENT, WAS UTILIZED UNDER IDE (B)(4); (SUBJECT OF THE QUEST CLINICAL STUDY) AND IS NOT THE APPROVED ALTIUS SYSTEM NERVE CUFF ELECTRODE (P230020/08-26-2024). NEUROS IS SUBMITTING THIS REPORT IN COMPLIANCE WITH FDA MEDICAL DEVICE REPORTING REGULATIONS UNDER 21 CFR PART 803.

Description of Event or Problem · 0

DURING 31-010'S IPG UPGRADE VISIT, SHE MENTIONED EXPERIENCING SHOCKING SENSATIONS A HANDFUL OF TIMES. SHE STATED THAT THEY BEGAN NOT LONG AFTER SHE FELL DOWN IN (B)(6) 2024 (THIS AE WAS REPORTED.) THE FALL RESULTED IN A BRUISE THAT STARTED FROM HER BACK AND ENDED RIGHT NEXT TO THE IPG. THE SHOCKING SENSATIONS ORIGINATED AT OR NEAR THE IPG AND TRAVEL DOWN HER STUMP. AS PART OF THE IPG UPGRADE VISIT, THE NEUROS FCS MEASURED THE DEVICE'S IMPEDANCE AND PERFORMED THRESHOLD TESTING. IMPEDANCE VALUES FLUCTUATED FROM 400 TO HIGH IMPEDANCE, BUT THEY WERE CONSISTENT WITH THOSE FROM PREVIOUS VISITS. THRESHOLD TESTING WAS ALSO CONSISTENT WITH PREVIOUS VISITS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1431358 ALTIUS SYSTEM NERVE CUFF ELECTRODE LGW NEUROS MEDICAL INC. FG-602-003

Patients

Seq Age Sex Outcome Treatment
1 64 YR Female Other