COLLEAGUE
Report
- Report Number
- 6000001-2011-21958
- Event Type
- Malfunction
- Date Received
- September 1, 2011
- Date of Event
- August 1, 2011
- Report Date
- August 16, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH DAMAGED BATTERY WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY DEPLETED MAIN BATTERIES DUE TO USE/USER ERROR. THE MAIN BATTERIES WERE REPLACED TO CORRECT THE REPORTED CONDITION. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 5.09.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROCESS.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION OF A DAMAGED BATTERY. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THE EVENT WAS REPORTED TO HAVE OCCURRED UPON POWER UP. THERE WAS NO REPORT OF PATIENT/USER INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |