FDA Adverse Event Malfunction Summary report: N

COLLEAGUE

MDR report key: 2230219 · Received September 1, 2011

Report

Report Number
6000001-2011-21942
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 1, 2011
Report Date
August 16, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER PERSONNEL, AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH MALFUNCTION OF WILL NOT PUMP BECAUSE OF FAILED CHANNEL WAS CONFIRMED BY BAXTER QUALITY ENGINEERING AS FAILURE CODE 810:11. THIS CONDITION WAS CAUSED BY THE AIR IN LINE PRINTED CIRCUIT BOARD (AIL PCB) BEING OUT OF CALIBRATION. THE AIL PCB WAS RE-CALIBRATED TO CORRECT THIS CONDITION. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE HAS NOT BEEN SERVICED FOR THE REPORTED CONDITION PRIOR TO THIS EVENT. A DEVICE HISTORY REVIEW WAS PERFORMED FINDING NO EXCEPTION, NONCONFORMANCE, OR REWORK THAT OCCURRED DURING THE MANUFACTURING OF THE COMPLAINT LOT OR SERIAL NUMBER. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A MALFUNCTION THAT THE DEVICE "WILL NOT PUMP BECAUSE OF FAILED CHANNEL". THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THE EVENT WAS REPORTED TO HAVE OCCURRED UPON POWER UP IN THE BIOMED SERVICE DEPARTMENT. THERE WAS NO REPORT OF PATIENT/USER INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.