FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2230161 · Received September 1, 2011

Report

Report Number
2050012-2011-04778
Event Type
Malfunction
Date Received
September 1, 2011
Date of Event
August 1, 2011
Report Date
August 1, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JGS
PMA / PMN Number
K042291
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

QC AT THE TIME OF THE EVENT WAS SHOWING IMPRECISION AND HIGH RECOVERY.THE CUSTOMER FOUND THAT THE MODULAR CHEMISTRY (MC) SAMPLE PROBE WAS PLUGGED WITH A PIECE OF RED RUBBER. CLEANING THE PROBE RESOLVED THE ISSUE.A FIELD SERVICE ENGINEER (FSE) WENT ON-SITE, EVALUATED THE INSTRUMENT AND VERIFIED PERFORMANCE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER GENERATED FALSE HIGH SODIUM (NA) AND CHLORIDE (CL) RESULTS. THE RESULTS WERE NOT REPORTED OUTSIDE OF THE LABORATORY. WHEN QC IDENTIFIED A PROBLEM, APPROXIMATELY 130 SAMPLES WERE RERUN ON A DIFFERENT INSTRUMENT IN THE LABORATORY AND THOSE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE RESULTS PROVIDED BY THE CUSTOMER ARE SHOWN IN THE ATTACHMENT. THERE WAS NO AFFECT TO PATIENT WITH REGARD TO THIS EVENT.