FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2230044 · Received August 31, 2011

Report

Report Number
2531779-2011-06395
Event Type
Malfunction
Date Received
August 31, 2011
Report Date
August 4, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: OBSERVATION CONFIRMED NO DAMAGE TO THE KEYPAD. THE OK AND DOWN BUTTONS WERE FOUND TO BE NOT RESPONDING APPROPRIATELY TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS AND THE OK BUTTON CONTACT WAS INVERTED.

Description of Event or Problem · 1

THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE ALLEGED KEYPAD ISSUE. REPORTEDLY, THE OK BUTTON REQUIRES SEVERAL PRESSES TO ENGAGE A RESPONSE. THERE WAS NO EVIDENCE OF PRODUCT MISUSE. THE PUMP IS BEING RETURNED FOR INVESTIGATION. THE PATIENT WILL GO ON THE BACKUP PLAN AS NEEDED. THERE WAS NO REPORT OF PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 62 YR