ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-06394
- Event Type
- Malfunction
- Date Received
- August 31, 2011
- Report Date
- August 4, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE RESULTS OF THE INVESTIGATION WERE AS NOTED: THE ISSUE WITH THE KEYPAD BUTTON MALFUNCTION WAS CONFIRMED THROUGH PRODUCT ANALYSIS. THE KEYPAD BUTTONS ARE INTERMITTENTLY RESPONDING TO USER INPUTS DURING TESTING. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER THE KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE ALLEGED KEYPAD ISSUE. REPORTEDLY, ALL THE KEYPAD BUTTONS ARE NOT RESPONDING TO USER INPUT. THERE WAS NO EVIDENCE OF PRODUCT MISUSE. THE PUMP IS BEING RETURNED FOR INVESTIGATION. THE PATIENT WILL GO ON THE BACKUP PLAN AS NEEDED. THERE WAS NO REPORT OF PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR |