FDA Adverse Event Injury Summary report: N

IMP,TSV,3.7,11.5,MTX,MG

MDR report key: 22293472 · Received June 19, 2025

Report

Report Number
0002023141-2025-01635
Event Type
Injury
Date Received
June 19, 2025
Date of Event
May 22, 2025
Report Date
September 11, 2025
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019935
PMA / PMN Number
K101977
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). IT IS UNKNOWN IF THE DEVICE WILL BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

ZIMVIE COMPLAINT NUMBER (B)(4). THE FOLLOWING FIELDS HAVE BEEN UPDATED: B4: DATE OF THIS REPORT. B5: DESCRIBE EVENT OR PROBLEM. G3: DATE RECEIVED BY MANUFACTURER . G6: TYPE OF REPORT. H1: TYPE OF REPORTABLE EVENT. H2: FOLLOW UP TYPE. H3: DEVICE EVALUATED BY MANUFACTURER. H6: ADVERSE EVENT PROBLEM. H10: ADDITIONAL NARRATIVE. ZIMVIE DID NOT RECEIVE ONE (1) TSVTB11, (IMP, TSV,3.7,11.5,MTX,MG) FOR EVALUATION. VISUAL INSPECTION COULD NOT BE PERFORMED. THE INVESTIGATION HAS BEEN PERFORMED BASED ON THE AVAILABLE INFORMATION USING THE ITEM NUMBER, DEVICE HISTORY RECORD (DHR) REVIEW, AND RISK MANAGEMENT FILE (RMF). DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1233255. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT, WERE NOTED AS PART OF THE DHR. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1233255 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. REVIEW COMPLETED UTILIZING KEYWORDS: DENTAL : FUNCTIONAL : FRACTURE : IMPLANT. THE CUSTOMER DID SUBMIT IMAGES FOR THE REPORTED EVENT. THE IMPLANT WAS FRACTURED AT THE COLLAR. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSE DETERMINED FROM THE INVESTIGATION WAS PATIENT FACTORS. THEREFORE, BASED ON THE PICTORIAL EVIDENCE AND ALL AVAILABLE INFORMATION, A DEVICE MALFUNCTION DID OCCUR. THE IMPLANT WAS FRACTURED AT THE TIP. THE REPORTED EVENT WAS CONFIRMED. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCTS WERE NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED. AT THIS TIME, THE COMPLAINT INVESTIGATION HAS BEEN COMPLETED AND THE RECORD WILL BE CLOSED. IF ADDITIONAL INFORMATION IS RECEIVED, THE RECORD WILL BE RE-OPENED FOR FURTHER EVALUATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE IMPLANT AT TOOTH SITE #19 FRACTURED. IMPLANT CROWN WAS LOOSE. RADIOGRAPH SHOWS FRACTURED ON DISTAL OF IMPLANT BODY. NEED TO REMOVE AND REPLACE IMPLANT AND CROWN.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
724532 IMP,TSV,3.7,11.5,MTX,MG DENTAL IMPLANT DZE ZIMMER DENTAL 1233255 00889024019935

Patients

Seq Age Sex Outcome Treatment
1 63 YR Unknown Required Intervention